Question 61

Which of the following is the standard administered activity range for an adult Tc-99m MDP bone scan?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: The usual adult dose of Tc-99m MDP or HDP is 20–30 mCi (740–1110 MBq). Pediatric doses are scaled by body weight or surface area. Higher activities increase radiation dose without proportional gain in diagnostic information. Technologists must assay the dose accurately, ensure proper hydration after injection, and schedule imaging 2–4 hours later to allow soft-tissue clearance. Knowledge of the standard dose range helps prevent dosing errors and supports ALARA practice while still providing adequate count density for high-quality planar and SPECT images.

Question 62

What is the most appropriate immediate action if a technologist’s ring dosimeter reading exceeds the institutional investigation level?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Investigation levels are set below regulatory limits so that higher-than-usual exposures can be reviewed before limits are approached. When a ring-badge reading exceeds the investigation level, the radiation safety officer must be notified. Together the RSO and technologist examine recent work habits, shielding use, time spent handling sources, and possible contamination of the badge. Corrective actions (additional shielding, practice changes, retraining) are implemented. Ring badges can be affected by geometry and contamination, but elevated readings are never ignored. Documentation of the investigation is required. Prompt attention prevents further unnecessary extremity dose and demonstrates a functioning ALARA program.

Question 63

Which of the following is the primary purpose of a geometry test on a dose calibrator?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Geometry testing evaluates how changes in sample volume, vial or syringe size, or position within the chamber affect the calibrator reading. Correction factors are established when significant differences are found. The test is performed at installation and after major repairs or when new container types are introduced. Daily constancy and quarterly linearity tests serve different purposes. Accurate geometry correction prevents systematic errors when assaying doses in containers different from those used for calibration. Technologists must apply the correct geometry factor when required by institutional protocol.

Question 64

Which of the following is the correct radiopharmaceutical for a Meckel’s diverticulum study?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Tc-99m sodium pertechnetate is taken up by gastric mucosa and is therefore the agent of choice for detecting Meckel’s diverticulum that contains ectopic gastric mucosa. Sequential imaging of the abdomen is performed for 30–60 minutes after injection. Premedication with H2 blockers, pentagastrin, or glucagon may be used according to institutional protocol to enhance uptake or reduce intestinal motility. Sulfur colloid, MDP, and MAA have different biodistributions and are not appropriate. Technologists must know the correct timing, positioning (including lateral views if needed), and possible pharmacologic interventions so that ectopic mucosa is not missed and so that physiologic gastric activity is correctly interpreted.

Question 65

In a three-phase bone scan, which phase is acquired immediately after injection and demonstrates blood flow?

Correct answer: C

Explanation

Comprehensive and Detailed Explanation: A three-phase bone scan consists of a dynamic flow (perfusion) phase acquired immediately after bolus injection of Tc-99m MDP or HDP, a blood-pool (soft-tissue) phase obtained a few minutes later, and delayed static or SPECT images acquired 2–4 hours post-injection. The dynamic flow phase shows arterial delivery of the tracer and is useful for evaluating vascularity of lesions such as osteomyelitis, reflex sympathetic dystrophy, or tumors. The blood-pool phase reflects soft-tissue hyperemia. Delayed images demonstrate osteoblastic activity. Proper timing, patient positioning, and framing rates are essential. Misidentifying the phases can lead to incorrect interpretation of infection versus other pathology. Nuclear medicine technologists must know the clinical indications that require a three-phase study versus a routine delayed bone scan and must ensure the radiopharmaceutical is injected as a rapid bolus for optimal flow images.

Question 66

Which of the following is the correct preparation requirement for a Tc-99m sestamibi kit that requires heating?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Most sestamibi kits require addition of Tc-99m pertechnetate followed by heating in a rolling boiling-water bath (or validated heating block) for a prescribed period (commonly 10 minutes) to complete the exchange reaction that forms the final complex. Incomplete heating results in poor radiochemical purity and excessive soft-tissue and liver uptake. After heating, the vial is cooled and the product is ready for clinical use within its beyond-use time. Technologists must follow the exact instructions for activity, volume, heating duration, and subsequent quality-control testing. Proper preparation is essential for high-quality cardiac and parathyroid imaging.

Question 67

What is the primary purpose of a radiation survey after a therapeutic I-131 administration before patient release?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Before a patient who has received therapeutic I-131 may be released, the residual dose rate at 1 meter is measured and compared with regulatory or institutional release limits. If the measured rate (or the calculated total effective dose to others) is below the applicable threshold, the patient may be released with written instructions. The survey is documented and retained. Technologists commonly perform or assist with this measurement. Accurate survey technique and proper interpretation of the result are essential for both regulatory compliance and protection of the public.

Question 68

Which interaction dominates for 140 keV photons in soft tissue?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: At 140 keV in soft tissue (effective Z ≈ 7.5), Compton scattering is the predominant interaction. The photoelectric effect becomes more important at lower energies or in higher-Z materials (bone, contrast agents, lead). Pair production requires photon energies above 1.022 MeV. Photodisintegration occurs only at very high energies. The dominance of Compton scatter explains why scatter correction and energy-window optimization are important in Tc-99m imaging: scattered photons lose energy and can still fall within a wide energy window, degrading contrast. Understanding the relative probabilities of these interactions also guides the design of collimators and the choice of attenuation-correction methods in SPECT and SPECT/CT.

Question 69

For a Tc-99m MDP bone scan, what is the typical recommended administered activity for an adult patient?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: The usual adult dose of Tc-99m medronate (MDP) or oxidronate (HDP) for a whole-body bone scan is 20–30 mCi (740–1110 MBq), with many laboratories using approximately 25 mCi. Lower activities may be used in pediatric patients (scaled by weight or body-surface area) or when SPECT/CT is the primary examination. Higher activities increase radiation dose without proportional diagnostic benefit. Proper hydration and frequent voiding after injection reduce bladder and radiation dose to the pelvic organs. Imaging is performed 2–4 hours post-injection to allow soft-tissue clearance. Technologists must verify the prescribed activity, assay the dose in a calibrated dose calibrator, and confirm patient identity and pregnancy status before administration. Dose ranges may be adjusted for specific clinical indications or institutional protocols, but 15–30 mCi remains the widely accepted adult range.

Question 70

What is the primary purpose of an ALARA investigation level?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Investigation levels are administrative thresholds (commonly 10–30 % of the regulatory limit) established by the licensee. When a dosimeter reading exceeds an investigation level, the radiation safety officer reviews the worker’s activities, technique, shielding use, and workload to identify the cause and implement corrective actions if needed. This proactive process supports ALARA by addressing elevated exposures before regulatory limits are approached. Investigation levels do not replace regulatory limits and do not automatically result in permanent restriction. Documentation of each investigation is retained as part of the radiation-protection program.

0%

Your score on this page: 0/0

Answer the questions above to see your score.

0 of 90 free Nuclear Medicine questions practiced · - overall accuracy

Want the full Nuclear Medicine question bank?

Unlock all 285 questions with mock exam, explanations for every option and domain-level analytics.

Page 7 of 9 · Questions 61-70 of 90 free

PracticeTestMaster is an independent exam prep provider. Nuclear Medicine is the property of American Registry of Radiologic Technologists (ARRT).

Full access

Unlock all 285 Nuclear Medicine questions

You're practicing 90 free questions. Paid plans add:

  • 195 more exam-style questions
  • Timed mock exam simulations
  • Domain-level score analytics
  • Unlimited reattempts
1 Month $19.99
Buy
12 Months $49.99
Buy
See everything included

Your Nuclear Medicine progress

Questions answered0
Accuracy-

Saved on this device.

Nuclear Medicine exam overview

Exam
Nuclear Medicine
Code
Nuclear Medicine
Credential body
American Registry of Radiologic Technologists (ARRT)
Updated
Nov 24, 2026
Unlock all 285 questionsFrom $19.99
Unlock