Nuclear Medicine Questions 31-40 of 90 free
0/10 answered · 0 correct

Choose an answer and click Check answer to see if you're right, with the full explanation.

Question 31

Which of the following is the primary advantage of a low-energy high-resolution (LEHR) collimator compared with a low-energy general-purpose (LEGP) collimator?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: LEHR collimators have smaller holes and longer septa than LEGP collimators, providing superior spatial resolution at the cost of reduced sensitivity. They are preferred when fine detail is important (e.g., bone SPECT, pediatric imaging, or cardiac SPECT when count rates permit). LEGP collimators are chosen when higher sensitivity is needed. Technologists must select the collimator appropriate for the clinical question and available activity. Using the wrong collimator can either degrade resolution or produce images with inadequate counts and increased noise.

Question 32

Which of the following is the correct sequence for preparing a typical Tc-99m MDP kit?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Most Tc-99m MDP (and HDP) kits are supplied as lyophilized vials containing the ligand and stannous reductant. The technologist simply adds the appropriate volume and activity of Tc-99m sodium pertechnetate and gently mixes. No boiling is required (unlike some sestamibi kits). Adding reagents in the wrong order or using excessive volume can reduce labeling efficiency. Package-insert instructions must be followed exactly regarding maximum activity, minimum and maximum volumes, and incubation time. After preparation, radiochemical purity testing may be required. Incorrect preparation can result in free pertechnetate (stomach and thyroid visualization) or colloidal impurities (liver and spleen uptake), both of which degrade bone-scan quality and increase non-target radiation dose.

Question 33

Which of the following is the primary interaction responsible for the photopeak in a NaI(Tl) spectrum?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: In the photoelectric effect the entire energy of the incident photon is absorbed by an orbital electron, resulting in full energy deposition in the crystal and appearance of the photopeak. Compton scattering deposits only part of the energy and produces the continuum below the photopeak. Pair production occurs only above 1.022 MeV. Understanding the origin of the photopeak allows technologists to set energy windows correctly, recognize the effects of scatter, and interpret changes in energy resolution that may indicate crystal or electronics problems.

Question 34

Which of the following is the approximate physical half-life of fluorine-18?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Fluorine-18 has a physical half-life of approximately 109–110 minutes. This short half-life requires efficient production, rapid transport, and careful scheduling of PET studies. Decay corrections must be applied accurately when calculating residual activity or comparing sequential scans. The half-life is long enough to permit synthesis of FDG and delivery to imaging centers within a reasonable radius of a cyclotron, yet short enough to limit patient radiation dose. Technologists working with PET must be proficient in rapid decay calculations and in coordinating injection and imaging times.

Question 35

What is the primary reason for restricting access to a high-radiation area?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: High-radiation areas are posted and access-controlled because an individual could receive a dose exceeding 1 mSv (100 mrem) in 1 hour at 30 cm from the source. Restricting entry, using interlocks or visual alarms when required, and limiting stay times are fundamental radiation-protection measures. Only trained personnel with appropriate dosimetry should enter, and the radiation safety officer must approve or supervise work in such areas. In nuclear medicine this situation most often arises after a major spill or during handling of high-activity therapy sources. Proper control prevents accidental overexposure and is a clear regulatory requirement.

Question 36

What is the primary purpose of an investigation level in a personnel dosimetry program?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Investigation levels are administrative thresholds (often 10–30 % of the regulatory limit) established by the licensee. When a dosimeter reading exceeds an investigation level, the radiation safety officer reviews the worker’s activities, technique, shielding use, and workload to determine the cause and to implement corrective actions if needed. This proactive process supports ALARA by identifying problems before regulatory limits are approached. Investigation levels do not replace regulatory limits and do not automatically result in restriction from work. Documentation of each investigation is retained as part of the radiation-protection program records.

Question 37

Which of the following is the correct action if a sealed source used for dose-calibrator constancy is found to be leaking?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Any sealed source that shows evidence of leakage (detectable removable contamination) must be immediately withdrawn from use, placed in a sealed container, and reported to the radiation safety officer. The RSO will arrange for proper testing, return to the manufacturer, or disposal as radioactive waste. Continuing to use a leaking source can spread contamination throughout the department and expose personnel. Regular wipe tests of sealed sources are required by regulation. Technologists must know the location of all sealed sources, perform or assist with leak testing according to schedule, and never attempt to open or repair a sealed source. Prompt reporting protects staff and maintains regulatory compliance.

Question 38

Which of the following is the correct elution solvent for a standard Mo-99/Tc-99m generator?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Mo-99/Tc-99m generators are eluted with sterile, pyrogen-free 0.9 % sodium chloride (normal saline). The saline removes Tc-99m pertechnetate from the alumina column while the parent Mo-99 remains bound. Other solvents can damage the column or introduce impurities. Technologists must use the volume recommended by the manufacturer, maintain aseptic technique, and perform required Mo-99 and aluminum breakthrough tests on the eluate. Proper elution technique maximizes yield and ensures the eluate is suitable for clinical use.

Question 39

Which of the following is the correct radiopharmaceutical for detection of Meckel’s diverticulum?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Tc-99m pertechnetate is actively taken up by gastric mucosa and is therefore the agent of choice for localizing Meckel’s diverticulum that contains ectopic gastric tissue. Sequential abdominal images are obtained for 30–60 minutes after injection. Pharmacologic premedication (H2 blockers, pentagastrin, or glucagon) may be used according to protocol to enhance uptake or reduce motility. Sulfur colloid, MDP, and MAA have different biodistributions and are not appropriate. Technologists must know the correct timing, positioning, and possible interventions so that ectopic mucosa is reliably detected and physiologic gastric activity is correctly interpreted.

Question 40

Which of the following is the correct method to reduce the number of particles in a Tc-99m MAA dose for a patient with pulmonary hypertension?

Correct answer: A

Explanation

Comprehensive and Detailed Explanation: Commercial MAA kits contain a fixed number of particles. For patients with pulmonary hypertension, right-to-left shunt, or pediatric patients, the number of particles administered must be reduced (often to 100,000–200,000). This is accomplished by reconstituting the kit according to instructions and then diluting a portion of the suspension with saline so that the desired particle number is contained in the injected volume. The activity is adjusted accordingly. Increasing boiling time or adding stannous ion does not change particle number. Technologists must calculate both activity and particle number and document the dilution. Administering too many particles can further compromise an already limited pulmonary vascular bed.

0%

Your score on this page: 0/0

Answer the questions above to see your score.

0 of 90 free Nuclear Medicine questions practiced · - overall accuracy

Want the full Nuclear Medicine question bank?

Unlock all 285 questions with mock exam, explanations for every option and domain-level analytics.

Page 4 of 9 · Questions 31-40 of 90 free

PracticeTestMaster is an independent exam prep provider. Nuclear Medicine is the property of American Registry of Radiologic Technologists (ARRT).

Full access

Unlock all 285 Nuclear Medicine questions

You're practicing 90 free questions. Paid plans add:

  • 195 more exam-style questions
  • Timed mock exam simulations
  • Domain-level score analytics
  • Unlimited reattempts
1 Month $19.99
Buy
12 Months $49.99
Buy
See everything included

Your Nuclear Medicine progress

Questions answered0
Accuracy-

Saved on this device.

Nuclear Medicine exam overview

Exam
Nuclear Medicine
Code
Nuclear Medicine
Credential body
American Registry of Radiologic Technologists (ARRT)
Updated
Nov 24, 2026
Unlock all 285 questionsFrom $19.99
Unlock