Nuclear Medicine Questions 51-60 of 90 free
0/10 answered · 0 correct

Choose an answer and click Check answer to see if you're right, with the full explanation.

Question 51

Which of the following is the primary purpose of a daily flood-field uniformity test on a gamma camera?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Daily intrinsic or extrinsic flood images are acquired to verify that the camera’s response remains uniform across the field of view. A new cold or hot area, or a significant change in quantitative uniformity values, indicates a problem such as photomultiplier-tube failure or crystal damage. Continuing to use a non-uniform camera can produce artifacts that mimic pathology. The flood is inspected visually and quantitatively; results are documented. If the flood is unacceptable, the camera is removed from service until repaired. This daily check is one of the most important quality-control procedures performed by nuclear medicine technologists.

Question 52

Which of the following is the standard imaging time after injection of Tc-99m sulfur colloid for a liver-spleen scan?

Correct answer: C

Explanation

Comprehensive and Detailed Explanation: After intravenous injection of Tc-99m sulfur colloid, imaging is typically begun 10–15 minutes later, once the particles have been cleared from the blood by the reticuloendothelial system. Earlier imaging may show residual blood-pool activity; much later imaging is unnecessary because the distribution is stable. Multiple views (anterior, posterior, laterals, and obliques) are obtained, and SPECT may be added. Technologists must ensure that the patient is properly positioned and that the count density is adequate. Recognition of normal variants (e.g., colloid shift to spleen and bone marrow in portal hypertension) and technical artifacts is essential for high-quality studies.

Question 53

Which of the following is the most likely cause of a sudden increase in background counts on a well counter?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: An abrupt rise in background counts most often indicates radioactive contamination inside or near the well. Residual activity from previous samples, a contaminated tube, or external contamination of the detector can all elevate background. Background must be measured and subtracted for accurate sample counting (e.g., thyroid uptakes, wipe tests, or blood volumes). If background is elevated, the well should be cleaned, resurveyed, and rechecked before further measurements. Technologists must count background regularly and maintain a clean counting area. Ignoring elevated background introduces systematic errors into quantitative results and may mask true contamination on wipe tests.

Question 54

What is the primary purpose of a COR (center-of-rotation) calibration on a SPECT system?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Center-of-rotation calibration determines any offset between the mechanical axis of the gantry and the electronic center of the detector matrix. Even small misalignments produce blurring and circular artifacts in reconstructed SPECT images. The calibration is performed with a point or line source and is required at regular intervals or after service. The resulting correction is applied during reconstruction. Uniformity maps, energy peaking, and decay correction are separate processes. Technologists must verify that the appropriate COR file is selected for the collimator and orbit type being used. Failure to maintain accurate COR calibration is a common source of preventable SPECT artifacts.

Question 55

Which of the following is the preferred radiopharmaceutical for renal cortical imaging to detect scarring?

Correct answer: C

Explanation

Comprehensive and Detailed Explanation: Tc-99m DMSA binds to the proximal tubular cells of the renal cortex and provides high-resolution images of cortical structure. It is the agent of choice for detecting pyelonephritic scarring, evaluating differential function, and assessing congenital anomalies. DTPA and MAG3 are cleared more rapidly and are preferred for dynamic functional studies. Imaging is performed 2–4 hours after injection, often with SPECT. Technologists must ensure adequate hydration, correct timing, and high-resolution collimation. Accurate DMSA imaging directly influences clinical decisions regarding antibiotic prophylaxis or surgical intervention in children with recurrent urinary-tract infections.

Question 56

Which of the following is the most appropriate radiopharmaceutical for detection of acute osteomyelitis in a three-phase bone scan?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: Tc-99m MDP (or HDP) is the standard agent for three-phase bone scintigraphy used to evaluate suspected acute osteomyelitis. Hyperperfusion and hyperemia in the flow and blood-pool phases combined with increased uptake on delayed images support the diagnosis, especially when radiographic changes are absent. Sulfur colloid is used for bone-marrow imaging or liver-spleen studies. Ga-67 and labeled leukocytes are reserved for more complex or chronic infection cases or when the bone scan is equivocal. Technologists must acquire true dynamic flow images (rapid sequential frames), a blood-pool image within a few minutes, and delayed images at 2–4 hours. Proper injection technique (bolus) and positioning are critical for diagnostic quality.

Question 57

What is the primary purpose of posting a “Caution – Radioactive Materials” sign?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: A “Caution – Radioactive Materials” sign is posted at entrances to rooms or areas where radioactive materials are used or stored in quantities exceeding regulatory thresholds. The sign alerts personnel and visitors that radioactive materials are present so that appropriate precautions can be taken. It does not by itself indicate a high-radiation area (which requires a different posting when dose rates exceed 100 mrem/hr at 30 cm). The sign does not eliminate the need for dosimetry or training. Proper posting is a basic regulatory requirement and is checked during inspections. Technologists must ensure that signs are visible, accurate, and removed when no longer applicable.

Question 58

Which of the following is the primary purpose of a bar-phantom resolution test?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: A four-quadrant bar phantom or orthogonal-hole phantom is imaged to assess spatial resolution (the smallest bars that can be clearly distinguished) and spatial linearity (straightness of the bars). The test is performed intrinsically or extrinsically at regular intervals. Loss of resolution or appearance of distortion indicates problems with the crystal, PMTs, or electronics. Energy resolution requires a pulse-height spectrum; dose-calibrator testing is separate. Routine resolution testing helps ensure the camera can produce diagnostically adequate images before patient studies are performed.

Question 59

Which of the following is the primary purpose of a well-counter energy calibration or peaking procedure?

Correct answer: A

Explanation

Comprehensive and Detailed Explanation: Peaking a well counter (or gamma camera) centers the pulse-height window on the photopeak so that the maximum number of primary photons is accepted while scatter and background are minimized. Incorrect peaking reduces counting efficiency and can introduce errors in quantitative measurements such as thyroid uptake, wipe tests, or blood-volume calculations. The procedure should be performed whenever a different radionuclide is counted or when instrument stability is in question. Technologists must verify the correct isotope setting and window width before acquiring patient or quality-control counts.

Question 60

Which of the following is the correct radiopharmaceutical for a gastric-emptying study of solids?

Correct answer: B

Explanation

Comprehensive and Detailed Explanation: The standardized solid-meal gastric-emptying protocol uses Tc-99m sulfur colloid thoroughly mixed into scrambled egg whites or a whole-egg sandwich so that the tracer remains bound to the solid phase. Liquid emptying is studied with In-111 DTPA or Tc-99m DTPA in water. MAA and MDP are not appropriate. Technologists must prepare the meal exactly according to the protocol, ensure the patient consumes it within the required time, and acquire images at the prescribed intervals (typically up to 4 hours). Accurate quantification of the percentage of meal retained depends on consistent technique and proper decay correction.

0%

Your score on this page: 0/0

Answer the questions above to see your score.

0 of 90 free Nuclear Medicine questions practiced · - overall accuracy

Want the full Nuclear Medicine question bank?

Unlock all 285 questions with mock exam, explanations for every option and domain-level analytics.

Page 6 of 9 · Questions 51-60 of 90 free

PracticeTestMaster is an independent exam prep provider. Nuclear Medicine is the property of American Registry of Radiologic Technologists (ARRT).

Full access

Unlock all 285 Nuclear Medicine questions

You're practicing 90 free questions. Paid plans add:

  • 195 more exam-style questions
  • Timed mock exam simulations
  • Domain-level score analytics
  • Unlimited reattempts
1 Month $19.99
Buy
12 Months $49.99
Buy
See everything included

Your Nuclear Medicine progress

Questions answered0
Accuracy-

Saved on this device.

Nuclear Medicine exam overview

Exam
Nuclear Medicine
Code
Nuclear Medicine
Credential body
American Registry of Radiologic Technologists (ARRT)
Updated
Nov 24, 2026
Unlock all 285 questionsFrom $19.99
Unlock