Question 41
Which of the following is the most appropriate frequency for performing an accuracy test on a dose calibrator?
Explanation
Comprehensive and Detailed Explanation: Accuracy testing compares the dose calibrator’s reading of calibrated reference sources with their known activities and is performed at installation and at least annually thereafter (or more frequently if required by the manufacturer or license conditions). Daily constancy checks monitor short-term stability; linearity and geometry tests are typically quarterly. Failure of an accuracy test indicates that the instrument may be giving systematically incorrect readings and must be repaired or recalibrated before further clinical use. Documentation of annual accuracy results is required for regulatory compliance.